EN
24
PREPARATION
- If you’ve used the bone grafting kit, perform
the CLEAN function with the tube-to-tube
connector inserted between the peristaltic
pump tube and the bone grafting kit tube
(see Chapter 02)
- Disconnect the tube-to-tube connector from
the two tubes (see Chapter 03)
MATERIALS NECESSARY
- Enzymatic detergent with pH 6-9
- Water
- Ultrasonic container
- Brush with soft nylon bristles
- Distilled water
- Compressed air
- Single-use sterilization packets
- Steam sterilization
CAUTION: Do not use hydrogen
peroxide. If you intend to disinfect the
connector, do not use hydrogen peroxide,
but only disinfectants with neutral pH
(pH7); always rinse with sterile water to
preserve the disinfection.
CAUTION: Once used, correctly dispose
of the enzymatic detergent solution, do
not recycle.
05.7
Û
CLEANING AND STERILIZATION OF THE TUBE-TO-TUBE
CONNECTOR (Optional)
CLEANING AND DISINFECTION METHOD - Tube-to-tube connector
STERILIZATION METHOD - Peristaltic pump tubes
The sterilization process in steam autoclave
must be performed setting the parameters
indicated below, which are validated by
Mectron and reach the sterilization safety
levels (SAL 10
-6
):
- 3 times Pre-vacuum
(minimum pressure 60 mBar).
- Sterilization temperature 132°C
(interval 0°C ÷ +3°C).
- Sterilization time 4 minutes.
- Minimum drying time 20 minutes.
All the sterilization phases must be performed
by the operator in compliance to the norms
UNI EN ISO 17665-1:2007, UNI EN ISO
556-1:2002 e ANSI/AAMI ST 46:2002.
WARNING: Checking infections - Parts
that can be sterilized - Diligently remove
all residues of organic dirt before the
sterilization.
CAUTION: Perform the sterilization
using a water steam autoclave only. Do not
use any other sterilization procedure (dry
heat, irradiation, ethylene oxide, gas, low
temperature plasma, etc.).
CAUTION: Do not exceed the allowed
load of the steam sterilizer.
1
Prepare an enzymatic detergent solution
with pH 6-9, as per the instructions of the
manufacturer;