CLEANING AND STERILIZATION
Place the connector in the ultrasound tank
submerged by the enzymatic detergent
solution at 40°C, for at least 10 minutes;
2
3
4
5
6
Remove the connector from the ultrasonic
tank, brush it with the brush with soft nylon
bristles, and rinse it with distilled water to
eliminate all residues of detergent;
Accurately rinse externally and internally,
blowing compressed air onto it;
Once you’ve finished the cleaning
operations, check the connector under an
adequate source of light, paying attention
for dirt residues and, if need be, repeat the
cleaning cycle;
Individually seal the connector in a single-
use sterilization packet.
Sterilize the connector in the steam
autoclave.