CLEANING AND STERILIZATION
05.4
Û
CLEANING AND STERILIZATION OF THE INSERTS
PREPARATION
- Disconnect the insert from the handpiece
with the torque wrench (see Chapter 03)
CAUTION: Always disconnect the insert
from the handpiece before proceeding to
its cleaning and sterilization.
MATERIALS NECESSARY
- Enzymatic detergent with pH 6-9
- Water
- Container for immersion in enzymatic
solution
- Brush with soft nylon bristles
- Compressed air
- Ultrasound tank
- Distilled water
- Single-use sterilization packets
- Steam sterilization
- Syringe
CAUTION: Do not use hydrogen
peroxide. If you intend to disinfect the
inserts, do not use hydrogen peroxide, but
only disinfectants with neutral pH (pH7);
always rinse with sterile water to preserve
the disinfection.
CAUTION: Once used, correctly dispose
of the enzymatic detergent solution, do
not recycle.
1
2
Prepare an enzymatic* detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
Place the insert horizontally in a clean
container, and add a quantity of enzymatic
detergent solution sufficient to completely
cover the insert;
* Process validated by independent institute with
enzymatic detergent Cidezime® (Enzol®).
CLEANING AND DISINFECTION METHOD - Inserts