STERILIZATION METHOD - Tube-to-tube connector
The sterilization process in steam autoclave
must be performed setting the parameters
indicated below, which are validated by
Mectron and reach the sterilization safety
levels (SAL 10
-6
):
- 3 times Pre-vacuum
(minimum pressure 60 mBar).
- Sterilization temperature 132°C
(interval 0°C ÷ +3°C).
- Sterilization time 4 minutes.
- Minimum drying time 10 minutes.
All the sterilization phases must be performed
by the operator in compliance to the norms
UNI EN ISO 17665-1:2007, UNI EN ISO
556-1:2002 e ANSI/AAMI ST 46:2002.
WARNING: Checking infections - Parts
that can be sterilized - Diligently remove
all residues of organic dirt before the
sterilization.
CAUTION: Perform the sterilization
using a water steam autoclave only. Do not
use any other sterilization procedure (dry
heat, irradiation, ethylene oxide, gas, low
temperature plasma, etc.).
CAUTION: Do not exceed the allowed
load of the steam sterilizer.
PREPARATION
- Disconnect the irrigation bag support rod
from the device body
- Disconnect the fixed support of the
handpiece from the device body
MATERIALS NECESSARY
- Enzymatic detergent with pH 6-9
- Container for solution with enzymatic liquid
- Clean and soft cloth with low
fiber-release
- Distilled water
- Single-use sterilization packets
- Steam sterilization
CAUTION: Once used, correctly dispose
of the enzymatic detergent solution, do
not recycle.
05.8
Û
CLEANING AND STERILIZATION OF THE IRRIGATION BAG
SUPPORT ROD AND OF THE HANDPIECE FIXED SUPPORT
CLEANING AND DISINFECTION METHOD
Irrigation bag support rod and handpiece fixed support
Prepare an enzymatic detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
1