PREPARATION
- Perform the CLEAN function
(see Chapter 02)
- Disconnect the handpiece from the device
body (see Chapter 03)
- Disconnect the peristaltic pump
tube from the connector of the handpiece
(see Chapter 03)
- If present, disconnect the insert from the
handpiece with the torque wrench (see
Chapter 03)
- Unscrew the front terminal
CAUTION: The handpiece and the cord
cannot be separated.
CAUTION: Disconnect the handpiece
from the device exclusively acting on the
connector.
CAUTION: Do not immerge the
handpiece in disinfectant solutions
or other liquids, because it could get
damaged.
CAUTION: Do not immerge the
handpiece in an ultrasonic tank.
CLEANING AND DISINFECTION METHOD - Handpiece with cord
Prepare an enzymatic* detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
* Process validated by an independent institute with
enzymatic detergent Cidezime® (Enzol®).
05.3
Û
CLEANING AND STERILIZATION OF THE HANDPIECE
1
MATERIALS NECESSARY
- Detergent solution (pH 6-9) and, if need be,
non-aggressive disinfectant solution with
neutral pH (pH7)
- Enzymatic detergent with pH 6-9
- Water
- Container for solution with enzymatic
liquid
- Clean, soft cloths with low fiber-release
- Brush with soft nylon bristles
- Distilled water
- Compressed air
- Single-use sterilization packets
- Steam sterilization
CAUTION: If you intend to disinfect,
we recommend that you use water-based
disinfectant solutions with a neutral pH
(pH7). Alcohol-based disinfectant solutions
and hydrogen peroxide are contra-
indicated, because they can fade the color
and/or damage the plastic materials. This
also holds true for chemical products such
as acetone and alcohol. Always rinse with
sterile water to preserve the disinfection.
CAUTION: Once used, correctly dispose
of the enzymatic detergent solution, do
not recycle.