CLEANING AND STERILIZATION
Delicately clean the surface of the
handpiece using the enzymatic detergent
solution with the aid of a brush with soft
nylon bristles, placing special care to the
following areas:
- threading of the handpiece
- titanium stem
- front terminal in its external and internal
parts
2
3
4
5
Accurately rinse with tap water to eliminate
any detergent residues and perform the last
rinse with distilled water;
Thoroughly dry all the parts, especially
the electrical contacts, blowing on them
with compressed air;
Clean the surface of the handpiece and of
its cord with a clean, soft cloth with low
fiber-release, dampened with a detergent
solution (pH 6-9) and, if need be, disinfect
with a non-aggressive disinfectant solution
with neutral pH (pH7), following the
instructions provided by the manufacturer
of the solution;