EN
4
Prepare an enzymatic detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
Immerge the end part of the peristaltic
pump tube or the tube of the bone grafting
kit, if used, in a container containing the
enzymatic detergent solution prepared;
1
2
3
4
Disconnect the flow adjuster from the
irrigation bag and from the peristaltic pump
tube or from the tube of the bone grafting
kit, if used.
Dispose of the flow adjuster correctly,
following the norms in force locally on the
disposal of hospital wastes;
Close the flow of the adjuster;
The CLEAN function allows to run a cleaning cycle on the irrigation circuit of the handpiece(s)
used during the treatment, by following the steps described in this chapter.
02
Û
CLEAN FUNCTION
CAUTION: Clean function. The CLEAN
function must be used after every
treatment, before starting the cleaning and
sterilization procedures.
CAUTION: If the tubes are not cleaned,
the crystallization of the salts may seriously
damage the device.
CAUTION: The handpiece and the cord
cannot be separated.