Leave to soak for 10 minutes at 40°C. This
procedure reduces the quantity of blood,
protein, and mucous present in the insert;
During immersion into the enzymatic
solution, delicately brush the surface of
the insert with a brush with soft nylon
bristles to eliminate all traces of visible dirt.
Accurately clean the difficult areas such as
sharp edges and especially the interstices
between the cutting cusps;
3
4
5
6
7
Delicately brush the surface of the insert
under tap water with the brush with soft
nylon bristles;
With a syringe, inject the enzymatic
detergent solution three times from inside
the cavity of the insert to effectively remove
the residues from the internal surface;
Rinse the cavity of the insert with distilled
water injected under pressure (3,8 bar) for
at least 10 seconds to eliminate all residues
of detergent;
x3
10 sec
3,8 bar