CLEANING AND STERILIZATION
Place the insert in the ultrasound tank
submerged by the enzymatic detergent
solution at 40°C, for at least 10 minutes;
Delicately brush the surface of the insert
again with the brush with soft nylon
bristles;
Remove the insert from the ultrasonic tank
and rinse with distilled water;
8
10
9
11
12
Rinse the internal canal of the insert
with distilled water injected under pressure
(3,8 bar) for at least 10 seconds to eliminate
all residues;
Once you’ve finished the cleaning
operations, check all the inserts under an
adequate source of light, paying attention
to the details that could conceal dirt
residues (threading, cavities, grooves) and, if
need be, repeat the cleaning cycle;
10 sec
3,8 bar