The 233v4 must be congured to suit the installaon and the ionising
equipment connected to it. Before you receive your 233v4, Meech, if
required, will preset the controller to suit your applicaon. This includes
applying basic sengs of the output power, rate and balance as well as alarm
funconality and the lock opon.
Note: If you are using it with a 971 ionising bar it will be preset to 15KV.
To change these sengs you will need a BarMaster or SmartControl Touch
and to follow the instrucons below.
Seng the Start Voltage and Ion Output Set Point
During installaon and commissioning, two parameters must be set correctly
to allow opmum performance:
1. Voltage Output. The correct start voltage must be set, depending on the
ioniser connected to the 233v4. The correct output voltages for Meech
ionisers are:
971 15 kV
242 10 kV
251 9 kV
261v2 9 kV
271/272/273 9 kV
976 Head 9 kV
976 Mini 9kV
2. Ion Output Set Point: This is the target Ion Output as measured when the
system is clean.