2. Specifications
Reference Conditions: 23°C ± 5°C (30 to 50% RH) range. Add 25 ppm/°C from
-10° to 18°C and 28° to 50°C to all accuracy specifications. No external electrical or
magnetic fields. Output current ≤150 mA for VT/PT and ≤ 50 mA for CT. Calibration
cycle is 1 year.
Electrical
Ratio Range (VT/PT) Autoranging: 0,8000 to 8000:1
Accuracy (VT/PT) Ratio range
Accuracy (% of reading)
0.8000 to 9.9999 ± 0.2%
10.000 to 999.99 ± 0.1%
1000.0 to 4999.9 ± 0.2%
5000.0 to 8000.0 ± 0.25%
Ratio Range (CT) Autoranging: 0.8000 to 1000.0
Accuracy (CT) Ratio Range
Accuracy (% of Reading)
0.8000 to 1000.0 ± 0.5%
Excitation signal Modo VT/PT: 32 V
RMS
máx.
CT mode: auto level 0 to 1 A, 0.1 to 4.5 V
RMS
Excitation current display Range: 0 to 1000 mA
Accuracy: ± (2% of reading + 2 mA)
Excitation Frequency 70 Hz
Display Dual line alpha-numeric LCD, 16x2 characters with
contrast adjustment and backlight control.
Day/night visible.
Measurement Method In accordance with IEEE Std C57.1290™-2006
Power Source Two 12 V, 5x2, 1650 mAH NiMH rechargeable
battery packs
Battery Life Up to 10 hrs of continuous operation. Low battery
indication.
Battery Charger
Universal input (90 to 264 Vrms input) smart recharger
Charging Time < 4 hrs
Data Storage 99 objects with 99 tests each
Date/ Time Battery-backed, Real-time clock
Communication USB. 2.0 compliant, optically isolated, 115.2 KB
Software Analysis software included