EasyManua.ls Logo

Megger Baker ADX - Page 41

Megger Baker ADX
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7. Ensure that all ADX test leads and Asset cables are properly separated from each other and safely run between the
Asset and the tester.
8. Press the power button on the lower left corner of the ADX.
Fig 35: Ensuring ADX test leads and Asset cables are properly separated and located.
CAUTION: Ensure that the ADX test leads are NOT coiled in any manner during testing to prevent
interference with test results. To avoid damaging the unit, DO NOT coil excess test leads and place
them on top of the ADX. Ensure that excess test lead cables are stretched out between/behind the
tester and Asset to ensure safe operation of both devices, and to reduce tripping and other hazards for
personnel in the area.
www.megger.com/baker
Megger Baker ADX Quick Reference Guide
35
Procedures for Testing with the ADX Analyzer

Table of Contents

Related product manuals