EasyManuals Logo

Megger MIT1525 User Manual

Megger MIT1525
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
27
Capacitance accuracy
(23 °C): ±10% ±5 nF
Voltage output accuracy
(>200V, 0 °C to 30 °C): +4%, -0%, ±10 V nominal test voltage
at 1 G
Current measurement
range: 0.01 nA to 6 mA
Current measurement
accuracy (23 °C): ±2% ±0.5 nA at all voltages
Interference (noise) rejection mA
rms
:
MIT515, MIT525: 1 mA per 250 V to a maximum of 3 mA
MIT1025: 1 mA per 600 V to a maximum of 3 mA
MIT1525: 1 mA per 350 V to a maximum of 6 mA
Voltmeter range: 30 V to 660 V ac or dc, 50/60 Hz
Voltmeter accuracy: ±3%, ±3 V
Timer range: Up to 99 minutes, 15 second minimum
setting
Memory capacity: hours continuous logging every 5 s.
or 33 logged PI tests, or 350 logged IR
tests
Test regimes:
MIT515 IR, IR(t), DAR, PI
MIT525, MIT1025: IR, IR(t), DAR, PI, SV, DD, ramp test
MIT1525: IR, IR(t), DAR, PI, SV, DD, ramp test
Interface: USB type B (device)
Real time output: USB, 1 reading/second (resistance,
current and voltage)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Megger MIT1525 and is the answer not in the manual?

Megger MIT1525 Specifications

General IconGeneral
BrandMegger
ModelMIT1525
CategoryTest Equipment
LanguageEnglish

Related product manuals