EasyManua.ls Logo

Megger MIT2500 - 3-Terminal Voltage Measurements

Megger MIT2500
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.megger.comMIT400/2
17
8.2 For 3 terminal instruments (MIT481/2, MIT485/2 only)
8.2.1 Using 2 test leads
1. Ensure instrument is in the A-B mode indicated in the display, as below:
- Figure 12
For details of changing the configuration see section 7.3 - 3 terminal connections.
2. Connect test leads to the RED/BLACK (B/A) test sockets only.
3. Select the VOLTAGE measurement mode on the range knob [
V
] range).
4. Connect the test leads to the circuit to be measured.
5. The instrument will display the voltage across the test leads as below.
100 0
8 00
6 004 00
2 00
0
A B
V
z
H
- Figure 13
Note: the display shows the A-B terminals are active.
6. If the display does not show the A-B option, press the A-E-B (T-G-R) button to change this mode, as below:
- Figure 14
7. With each press, the display will change from:
to to and back to
- Figure 15
Or, if 'T-G-R' is enabled in setup:

Table of Contents

Other manuals for Megger MIT2500

Related product manuals