EasyManua.ls Logo

Megger MIT2500 - Result Storage

Megger MIT2500
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.megger.comMIT400/2
18
to to
- Figure 16
8.2.2 Using 3 test leads
All three connections can be connected to the circuit under test, for example:
Electrical testing Telecommunications USA Telecommunications EU
A = Neutral T = Tip A = A
B = Live R = Ring B = B
E = Earth E = Earth E = E
The measurement should default to Live - Neutral (A-B) when the instrument is switched on.
Pressing the A-E-B (T-G-R) button will cycle through the voltages on each part of the circuit, see section 7.2.
8.3 Result storage
For MIT420/2, 430/2, 481/2, 485/2 and the MIT2500 the measured result may be stored (if required) by pressing the ‘STORE’
button for > 1sec. Refer to section 12 for further information.
WARNING - Where the A-E-B (T-G-R) symbol disappears, a voltage is present on a pair not currently selected, Press A-E-B
(T-G-R) to locate the live pair or pairs.

Table of Contents

Other manuals for Megger MIT2500

Related product manuals