www.megger.comMIT400/2
40
Frequency:
Frequency measurement range 15-400Hz
Frequency measurement accuracy ±0.5% ± 1digit (100 Hz to 400 Hz) unspecified
Capacitance measurment:
MIT420/2, MIT430/2
Capacitance measurement: 1 nF to 10 μF
Capacitance measurement accuracy ± 5.0% ± 2 digits
(0.1 - 1 nF unspecified)
Storage:
Result storage (MIT420/2 & MIT430/2):
Storage capacity >1000 test results
Data download Bluetooth
®
wireless
Bluetooth
®
Class II
Range up to 10 m
Power supply 6 x 1,5 V cells type IEC LR6 (AA, MN1500, HP7, AM3 R6HP)
6 x 1,2 V NiMH (HR6) rechargeable cells may be used
Battery life 3000 insulation tests with duty cycle of 5 sec ON /55 sec OFF @ 1000 V into 1 MW
Charger (Optional): 12-15 V dc (accessory interface)
Dimensions Instrument 228 mm x 108 mm x 63 mm
(9.00 in x 4.25 in x 2.32 in)
Weight 600 g (MIT400/2), (28.74 oz )
815 g (MIT2500) (27.22 oz)
Weight (instrument and case) 1.75 kg (3.86 lb)
Fuse Use only 500 mA (FF) 1000 V
(1 OR 2 depending on variant)
32 x 6 mm ceramic fuse of high breaking capacity HBC 30 kA minimum.
Glass fuses MUST NOT be fitted.
Safety protection The instruments meet EN 61010- 1 (1995) to 600V phase to earth, Category IV.
Refer to safety warnings supplied.
EMC In accordance with IEC 61326 including amendment No.1
Temperature co-efficient <0,1% per °C up to 1 GW
<0,1% per °C per GW above 1 GW
Environmental:
Operating temperature range and humidity -10 to +55 °C
90% RH at 40 °C max.
Storage temperature range -25 to +70 °C
Calibration temperature +20 °C
Maximum altitude 2000 m
IP rating IP 54