Service
AVTMTO210 Rev 7 Sept 2013
59
CODE Error Description Possible Cause
241 Onboard 10v out of range, voltage too low
242 Onboard 28v out of range, voltage too low
243
Discharge temperature is out of range, too
high
Unit overheated, fans filters clogged, fans
not working, overuse, extremely hot test
conditions
244
VICOR temperature is out of range, too
high
Unit overheated, fans filters clogged, fans
not working, overuse, extremely hot test
conditions
245 VNOISE voltage out of range, too low.
246 VICOR temperature too much ripple
3.0
ADC_ERROR
301 ADC reading times out or ADC locks up
302 Internal error. Communication with ADC
failed.
303 ADC sample rate setting failed
304 ADC internal zero cal failed
305 ADC internal full-scale calibration failed
306 ADC cannot be restored back to normal
mode
307 ADC channel 1 selection failed
308 ADC channel 2 selection failed
309 ADC channel 3 selection failed
310 Too much noise/ripple on ADC readings
311 ADC failed, initialization
312 ADC failed, run time
313 ADC locks up, can not read