52 D24030 | Date 11/2018
Menu mode
CAL 1
+0,5°C
CAL 3
+1,6°C
CAL 2
-0,4°C
Factory calibration
20°C 30°C 40°C
0°C
10°C
Fig. 28 Schematic example of temperature adjustment
Example: Temperature deviation at 30 °C should be corrected.
1. Press the activation key to the right of the
CALIB display. The display is enlarged
and the temperature adjustment option
is automatically highlighted.
Cal1
°C -0,2
K
Cal2 100.0 °C +0,1K
Cal3 180.0 °C -0,2K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
CK0,2-
CK0,1+
CK0,2-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
20.0
40.0
5.0
Temperature
Humidity
2. Press the confirmation key repeatedly,
until the calibration temperature Cal2 is
selected.
CK0,2-
CK0,1+
CK0,2-
Temperature
Cal1
Cal2
Cal3
40.0
5.0
20.0
3. With the turn control, set the calibration
temperature Cal2 to 30 °C.
CK0,2-
CK0,1+
CK0,2-
Temperature
Cal1
Cal2
Cal3
40.0
5.0
30.0
4. Save the setting by pressing the confir-
mation key. The corresponding calibra-
tion value is automatically highlighted.
CK0,2-
CK
CK
0,2-
Temperature
Cal1
Cal2
Cal3
40.0
5.0
30.0 0,1+