D24030 | Date 11/2018 53
Menu mode
5. Set the calibration value to 0.0 K and ac-
cept the setting by pressing the confirma-
tion key.
CK0,2-
CK
CK
0,2-
Temperature
Cal1
Cal2
Cal3
40.0
5.0
30.0 0,0
6. Position the sensor of a calibrated refer-
ence instrument centrally in the appli-
ance's working chamber.
7. Close the door and, in manual mode,
adjust the set temperature to 30 °C.
21.4°C
TEMP
30.0 °CSet
8. Wait until the appliance reaches the set
temperature and displays 30 °C. The
reference instrument for example displays
31.6 °C.
30.0°C
TEMP
Set 30
.0°C
31.6 °C
9. In the SETUP, adjust the calibration value
Cal2 to +1.6 K (actual value measured
minus setpoint temperature) and save the
setting by pressing the confirmation key.
CK0,2-
CK
CK
0,2-
Temperature
Cal1
Cal2
Cal3
40.0
5.0
30.0 1,6+
10. After the calibration procedure, the
temperature measured by the reference
instrument should now also be 30 °C.
30.0°C
TEMP
Set 30
.0°C
30,0 °C
With Cal1, a calibration temperature below Cal2 can be programmed accordingly, and with
Cal3, a temperature above. The minimum difference between the Cal values is 10 K.
If all calibration values are set to 0.0 K, the factory calibration settings are restored.