56 D39374 | Date 08/2018
Menu mode
5. Set the calibration value to 0.0 K and
accept the setting by pressing the con-
firmation key.
CK0,2-
CK
CK
0,2-
e
Cal1
Cal2
Cal3
180.0
40.0
120.0 0,0
6. Place the sensor of a calibrated refer-
ence meter on the thermoshelf to be
calibrated.
7. Close the door and, in manual mode,
adjust the set temperature to 120 °C.
ON
min
000°C
of °C
max
000°C
auto off
99K
-
+
GRAPH
76mb
Set 50 mb
In 1
°C100.2
TEMP
°CSet 100.0
100,4 °C 4
3
100,1 °C 2
100,2 °C 1
max
mb
min
16
TIMER
m4h
Ende
13:30 23.11.
Manueller Modus
12.09.2012 13:44
21.4°C
TEMP
Set
120.0 °C
8. Wait until the appliance reaches the set
temperature and displays 120 °C. The
reference instrument should display
122.6 °C.
120.0°C
TEMP
Set 120
.0°C
122.6 °C
9. In the SETUP, adjust the calibration
value Cal2 to +2.6 K (actual value
measured minus setpoint temperature)
and save the setting by pressing the
confirmation key.
CK0,2-
CK
CK
0,2-
re
Cal1
Cal2
Cal3
180.0
40.0
120.0 2,6+
10. After the calibration procedure, the
temperature measured by the reference
instrument should now also be 120 °C.
120.0°C
TEMP
Set 120
.0°C
120.0 °C
With Cal1, a calibration temperature below Cal2 can be programmed accordingly, and
with Cal3, a temperature above. The minimum clearance between the Cal values is 20 K.
If all calibration values are set to 0.0 K, the factory calibration settings are restored.