Guidelines on electromagnetic compatibility – EMC guidelines
This device was tested according to IEC 60601-1-2:2014 and meets all corresponding
requirements.
Portable HF (high frequency) communication devices, including antennas, can have a
negative influence on the device. Please operate the device (including accessories and
cables) with a minimum distance of 30 cm from such HF devices.
The main performance characteristic of this electrical medical device is defined as
application time with a maximum duration of 60 minutes. This main performance
characteristic can cannot be influenced by the negative impact of electromagnetic radiation.
Regarding the basic and EMC safety, no special maintenance measures are required for this
medical electrical device.
The “VM 9100RM III therapy system” is intended for operation in the electromagnetic
environment indicated below.
The operator has to ensure that the “VM 9100RM III therapy system” is operated in this
kind of environment.
Measurements of the emitted
electromagnetic interference
Electromagnetic
environment - Guidelines
Radiofrequency emissions acc. to
CISPR 11
The “VM 9100RM III
therapy system” is suitable
for use at all facilities,
including those in living
areas and at facilities which
are directly connected to a
public supply network
which also supplies
buildings which are used
for residential purposes.
Harmonic emissions acc. to
IEC 61000-3-2
Emission of voltage fluctuations and
flickers acc. to IEC 61000-3-2