5 MENUS AquaLab TDL
If you turn auto save on, all test readings save to the instruments
memory, but only the final reading appears on the main Measure-
ment screen. If you keep the TDL connected to a computer using
AquaLink 4 (See Section 11: Computer Interface), all readings taken
during a custom mode test download to the AquaLink 4 software.
On the Mode screen at the top of the page is the current mode
settings with the number of tests appearing first, followed by the sta-
bility value (∆a
w
). Pressing Enter with the custom mode highlighted
allows you to change the number of tests and stability settings.
To change the number of readings, use the right/left arrow buttons
to highlight the number under Readings, and then use the up and
down buttons to change to any value between 2 and 9.
To change the stability setting, use the right/left arrow buttons to
highlight the number under (∆a
w
), and then use the up and down
buttons to change to any value between 0.0005 and 0.0200. To save
the settings and finish, press the save button (to exit without updat-
20