15 APPENDIX A AquaLab TDL
15 Appendix A
15.1 Preparing Salt Solution
If you choose to mix a saturated salt solution for use as a verification
standard, we recommend that you use the approved AOAC method.
Steps 1 through 4 detail the AOAC method.
1. Select a reagent-grade salt and place it in a test container to
a depth of about 4 cm for more soluble salts (lower a
w
), to a
depth of about 1.5 cm for less soluble salts (high a
w
), and to
an intermediate depth for intermediate salts.
2. Add distilled water in increments of about 2 mL, stirring con-
stantly.
3. Add water until the salt can absorb no more water, evidenced
by the presence of free liquid. Keep the amount of free liquid to
the minimum needed to keep the solution saturated with water.
If you plan on using this solution over a long term period, seal
the solution well to prevent losses from evaporation. Table 4
shows saturated salt solutions and their respective water ac-
tivities at various temperatures. Please note that these values
are based on averaged published data, and the standard errors
shown reflect Greenspan’s standard error for each salt solution,
not the AquaLab accuracy in measuring the salt. AquaLab
TDL measures all samples with an accuracy of ±0.005 a
w
.
90

Other manuals for METER AQUALAB TDL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the METER AQUALAB TDL and is the answer not in the manual?

METER AQUALAB TDL Specifications

General IconGeneral
BrandMETER
ModelAQUALAB TDL
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals