EasyManua.ls Logo

METER ATMOS 41 - Page 31

METER ATMOS 41
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
ATMOS 41
a. Rain funnel (Figure16)
b. Pyranometer (Figure16)
c. Ultrasonic transducer openings (Figure13)
d. Sintered glass reflection plate (Figure7)
3. Clean the ATMOS41 body.
a. Scrub with light to medium pressure using a warm, damp cloth.
b. Clean around posts and between crevices using a dry brush.
CAUTIONS
DO NOT immerse the sensor in water.
DO NOT touch the temperature sensor needle (Figure13).
CAUTION: The temperature sensor needle lead wires are very delicate and can be easily damaged.
• Avoid more than light pressure on the sonic transducers (Figure13).
• Be sure the sensor is level after cleaning.
Temperature sensor
Ultrasonic
transducers
Ultrasonic
transducers
Figure13 Temperature needle and ultrasonic transducers
NOTE: Do not allow water to enter the ultrasonic sensors (Figure13). Water may corrode the metal parts inside the
sensors and ruin them. Do not touch the temperature sensor when cleaning because it is very delicate and can be
damaged if pushed into the ATMOS41 body.
c. Check the downspout (Figure14) for debris.
d. Observe the Teflon screen (Figure15) to see if it is dirty.
If the screen is dirty, it is best to replace it.
Contact Customer Support for a replacement Teflon screen.

Other manuals for METER ATMOS 41