Questions and Answers

  • J
    John WoodAug 20, 2025
    What to do if METER Weather Station ATMOS 41 is not reading any rain?
    • M
      Martin GomezAug 20, 2025
      First, inspect the rain gauge funnel and clear away any obstructions. Ensure the ATMOS 41 is leveled within approximately ±2 degrees in both the X and Y directions for accurate rainfall measurements. If it's not level, drops might miss the electrodes. Check level measurements in the ATMOS 41 data stream to confirm. Gently twist and remove the rain gauge funnel, being careful to unplug the pyranometer connector inside before removing the funnel completely. Inspect for any issues and ensure the gold electrodes are correctly aligned.
  • M
    miranda34Aug 22, 2025
    How to fix no wind speed on METER Weather Station?
    • J
      John SmithAug 22, 2025
      Check the anemometer pathway, ensuring nothing blocks the sonic transducer measurement path. Inspect the sonic transducers for water build-up; if present, gently dab it away with a dry cloth. Ensure the acoustic mirror is clean by flushing it with water and drying it with a cloth. Also, make sure the ATMOS 41 is level.
  • S
    Samuel GomezAug 25, 2025
    What to do if METER ATMOS 41 Weather Station is not responding?
    • M
      msanchezAug 25, 2025
      Check the sensor's power supply, cable, and the 3.5-mm plug. Verify the data logger wiring connections: Brown to 3.6 V to 15 V power, Orange to digital out, and Bare to ground. If the sensor remains unresponsive, use the ProCheck to ensure it is working correctly.
  • A
    Alexa HinesAug 26, 2025
    What to do if my METER ATMOS 41 Weather Station has no temperature reading?
    • M
      Michelle JuarezAug 26, 2025
      Check the temperature needle to ensure it hasn't been pushed in, as this can break the thermistor wires and stop the measurement. Be careful when cleaning around the temperature needle to avoid damaging its delicate lead wires.
  • P
    Pamela ElliottAug 27, 2025
    Why my METER ATMOS 41 pyranometer is not reading?
    • B
      Benjamin SalazarAug 27, 2025
      Carefully remove the rain funnel. Be careful to unplug the pyranometer connector inside the funnel before fully removing it. Ensure the pyranometer plug is properly connected.
  • L
    Leonard BarreraAug 29, 2025
    How to resolve water not flowing through rain gauge in METER ATMOS 41 Weather Station?
    • D
      dmartinAug 29, 2025
      Check all screens and the outflow to ensure there is no lodged debris.
  • H
    Heather BriggsAug 31, 2025
    Why is my METER ATMOS 41 Weather Station not reading rain accurately?
    • J
      Julie ScottAug 31, 2025
      If your METER Weather Station ATMOS 41 is not reading any rain or is measuring it inaccurately, first, clear any obstructions from the rain gauge funnel. Ensure the ATMOS 41 is within approximately ±2 degrees of dead level in both the X and Y directions. Use the internal level measurements in the ATMOS 41 data stream to confirm it's level. Gently twist the top of the weather station and remove the rain gauge funnel to inspect for any apparent issues and check that the gold electrodes are aligned correctly. You can also check the drip counter calibration using the procedure described in How to check the drip counter calibration (metergroup.com/atmos-41-drip-counter-calibration).
  • B
    Barbara BishopSep 2, 2025
    How to troubleshoot METER Weather Station ATMOS 41 not responding?
    • B
      brownbrianSep 2, 2025
      If your METER Weather Station ATMOS 41 is not responding, verify the power supply to the sensor. Inspect the sensor cable and stereo plug connector for any damage. Try using a different port on the data logger. Ensure the data logger wiring is correctly connected, paying attention to these connections: Brown—3.6- to 15.0-V power supply, Orange—digital out, and Bare—ground. If the sensor still doesn't respond, use the PROCHECK to ensure it's functioning correctly.
  • P
    Peter DuffySep 4, 2025
    What to do if my METER ATMOS 41 Weather Station shows no wind speed?
    • B
      Brandon EdwardsSep 4, 2025
      If your METER Weather Station shows no wind speed, check the anemometer pathway for any obstructions blocking the sonic transducer measurement. Inspect the sonic transducers for any water accumulation and gently dab it away with a dry cloth. Ensure the sintered glass plate is clean by flushing it with water and drying it with a cloth. Make sure the ATMOS 41 is level.
  • B
    bberrySep 6, 2025
    What to do if my METER ATMOS 41 has no pyranometer reading?
    • K
      Kristen SalazarSep 6, 2025
      If there is no pyranometer reading on your METER Weather Station, carefully remove the rain funnel. Ensure that the pyranometer plug is securely connected. If the pyranometer connector is damaged, contact Customer Support.

Summary

2. OPERATION

2.1 INSTALLATION

Steps to set up the ATMOS 41 and start collecting data.

2.2 CONNECTING LOGGERS

Connecting the ATMOS 41 to METER and third-party data loggers.

3. SYSTEM

3.1 MEASUREMENT SPECIFICATIONS

Detailed technical specifications for the ATMOS 41 sensors.

4. SERVICE

4.1 CALIBRATION

Details on how ATMOS 41 sensors are calibrated.

4.2 RECALIBRATION RECOMMENDATIONS

Recommended frequencies for sensor recalibration.

4.5 TROUBLESHOOTING

Guidance on diagnosing and resolving common issues.

METER ATMOS 41 Specifications

General IconGeneral
Measurement ParametersAir temperature, Relative humidity, Barometric pressure, Wind speed, Wind direction, Solar radiation, Precipitation
Wind Speed Range0 to 30 m/s
Wind Speed Accuracy±0.3 m/s or ±3% (whichever is greater)
Wind Direction Accuracy±3°
Solar Radiation Accuracy±5%
Relative Humidity Range0 to 100% RH
Relative Humidity Accuracy±2%
Barometric Pressure Accuracy±0.5 hPa
Wind Direction Range0° to 360°
Precipitation Accuracy±5%
Output InterfaceSDI-12
Operating Temperature-40°C to 60°C
CommunicationSDI-12
Air Temperature Accuracy±0.2°C (typical) from -40 to 60°C