Page #20 background imageLoading...
Page #20 background image
Functional description
20 100.000-count Graphical Multimeters
Functional description (cont'd)
SPEC
On the basis of the technical specifications, the SPEC mode directly displays
the tolerance of the measurement in progress, so that there is no need to
search for it and calculate it.
On the basis of the main measurement, the display:
indicates the specifications (x % of reading ± n digit) according to the type of
measurement, the range selected and the frequency (in AC and AC+DC)
calculates the interval containing the true value, if the instrument is within its
tolerance: SMIN value minimum specification
SMAX value maximum specification
Example:
MEM
MEM mode records the contents of the digital display(s) in the memory of the
instrument at a pre-programmed rate.
A short press on starts a series of recordings.
The MEM symbol flashes throughout the recording period; it is accompanied
by the number of recordings made.
Memorization of the measurements can be stopped by another short press.
The number of values to be memorized for a measurement run is
programmable: it therefore stops recording automatically.
Another press on starts a new series of recordings.
MTX 3281
MTX 3282, MTX 3283
158 measurements per sequence 1 to 10 sequences
Recording
capacity
6500 measurements maximum
1 to 10 sequences
(depending on available memory)
Example: activation
of MEM mode during
surveillance mode
Furthermore, the MEM
mode may be activated
during the SURV mode or
during display of the
secondary functions.
The parameters set
are saved.
It will then be possible to
select them and display
them as the main
function.
(When it is started up, the MTX 3281 reinitializes the date and time
(01:01:2000, 00 :00 :00).
Before starting work, set the correct date and time to "date-stamp" the
records. See §. Setting the date and time).
SPEC
1
.
0000
0.3%, ± 40 digits
SMIN
AC+DC
V
SMAX
1
.0070
V
0
.9930
V
0 5 10
In AC+DC, the
specifications are
calculated only if the
frequency can be
measured (see §.
secondary Functions)
and is > 45 Hz.
SURV
1
.
5000
1.3000
V
MIN
DC
V
MAX
AVG
2.5000
V
-
1.0000
V
-10 0 10
MEM 2
AUTO