EasyManua.ls Logo

Metrohm 813 - Appendix; Error Messages

Metrohm 813
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.1 Error Messages
30
813 Compact Autosampler
4 Appendix
4.1 Error Messages
An error occurring during a sample series is displayed by a blink-
ing red LED. The number of flashes indicates the error number. If
an error occurs during processing of a sample series, the
changer will then be switched into the 'HOLD' state. The current
method has to be halted by pressing <STOP>.
The list of possible error messages and their causes:
1* RAM defect
Call Metrohm-Service.
3* battery low
The battery for the permanent storage of the settings
must be replaced.
6* changer low power
The power supply cannot deliver enough power for the
simultaneous operation of all components currently in
use (pump and lift).
7* rack data missing
No sample rack is in position or no rack data can be
found for the sample rack that is in place.
8* invalid rack code
The rack code read by the changer could not be found
in the internal position tables.
10* raise lift first
Turning of a rack could not be carried out because the
lift was below the rest (shift) position.
12* changer overload
Load or resistance too large to turn the rack.

Table of Contents