4.2 Sequences
32
813 Compact Autosampler
Method 2
Single determination, no air gap, needle remains in sample after
sample transfer.
813 Compact Autosampler 813.0010
parameters
method 813_2
number of samples: *
>start sequence
1 SAMPLE: = 1
>sample sequence
1 SCN:Rm: *****1**
2 MOVE 1 : sample
3 LIFT: 1 : work mm
4 SCN:Rm: *****1**
5 PUMP 1.1 : on
6 WAIT 5 s
7 SCN:Rm : *****1**
8 WAIT 12 s
9 PUMP 1.1 : off
>final sequence
>changer settings
rack number 0
lift rate 12 mm/s
shift rate 20
shift direction: auto.
>manual stop
------------
¬ Program version
¬ Method name
¬ Number of samples unlimited
¬ First rack position
¬ Await 1
st
Remote signal
¬ Move next sample to tower
¬ Lift to work position / immerse needle
¬ Await 2
nd
Remote signal
¬ Switch on pump
¬ Waiting time
¬ Await 3
rd
Remote signal
¬ Waiting time for sample transfer
¬ Switch off pump
------ Settings for changer functions --------------