4.2 Sequences
34
813 Compact Autosampler
Method 4
Triple determination of each sample, no air gap, needle remains in
sample after sample transfer.
813 Compact Autosampler 813.0010
parameters
method 813_4
number of samples: *
>start sequence
1 SAMPLE: = 1
>sample sequence
1 SCN:Rm: *****1**
2 MOVE 1 : sample
3 LIFT: 1 : work mm
4 SCN:Rm: *****1**
5 PUMP 1.1 : on
6 WAIT 5 s
7 SCN:Rm : *****1**
8 WAIT 12 s
9 PUMP 1.1 : off
10 SCN:Rm: *****1**
11 WAIT 5 s
12 SCN:Rm: *****1**
13 PUMP 1.1 : on
14 WAIT 5 s
15 SCN:Rm : *****1**
16 WAIT 12 s
17 PUMP 1.1 : off
18 SCN:Rm: *****1**
19 WAIT 5 s
20 SCN:Rm: *****1**
21 PUMP 1.1 : on
22 WAIT 5 s
23 SCN:Rm : *****1**
24 WAIT 12 s
25 PUMP 1.1 : off
>final sequence
>changer settings
rack number 0
lift rate 12 mm/s
shift rate 20
shift direction: auto.
>manual stop
------------
¬ Program version
¬ Method name
¬ Number of samples unlimited
¬ First rack position
¬ Await 1
st
Remote signal
¬ Move next sample to tower
¬ Lift to work position / immerse needle
¬ Await 2
nd
Remote signal
¬ Switch on pump
¬ Waiting time
¬ Await 3
rd
Remote signal
¬ Waiting time for 1
st
sample transfer
¬ Switch off pump
¬ Await 1
st
Remote signal again
¬ Waiting time
¬ Await 2
nd
Remote signal
¬ Switch on pump
¬ Waiting time
¬ Await 3
rd
Remote signal
¬ Waiting time for 2
nd
sample transfer
¬ Switch off pump
¬ Await 1
st
Remote signal again
¬ Waiting time
¬ Await 2
nd
Remote signal
¬ Switch on pump
¬ Waiting time
¬ Await 3
rd
Remote signal
¬ Waiting time for 3
rd
sample transfer
¬ Switch off pump
------ Settings for changer functions --------------