4. ADV. ALG. (Advisory Algorithms) (F2).
These ECG algorithms are meant to test the analysis and prompt-
ing feature of automatic and semi-automatic defibrillators.
Choose desired selection by pressing UP (F2) or DOWN (F3).
Save this under ‘Wave” in the STATUS field by pressing
SELECT (F4). Press CANCEL (F5) to cancel selection.
5. CHARGE TIME (F3). Used to test the battery and charging
capacitor in the defibrillator. It changes the text ‘Delay’ to
‘Chrg T’ in the RESULT field in the main menu.
6. PRINT HEADER (F4). Automatically writes a heading for the
new test protocol.
7. WAVE AMPL. (Wave Amplitude) (F1).
Choose desired amplitude by pressing UP (F2) or DOWN (F3).
Save this under ‘Ampl” in the STATUS field by pressing
SELECT (F4). Press CANCEL (F5) to cancel selection.
8. PLAY PULSE (F2) enables playback of the last discharge.
9. PERF. WAVE (Performance ECG) (F3).
Choose desired wave by pressing UP (F2) or DOWN (F3). Save
this under ‘Wave” in the STATUS field by pressing SELECT
(F4). Press CANCEL (F5) to cancel selection.
10. SYSTEM TEST (F1) .
5