26
5 segments blinks: The electrical
connection to the right motor is de-
fective (short-circuit). Check the mo-
tor connection cable, motor cable
plug and motor.
6 segments blinks: The drive dis-
able function is active. The drive dis-
able function is an electronic security
function that prevents the wheelchair
from being driven when a battery
charger is connected.
7 segments blinks: A system or joy-
stick error. Movement of the joystick
during ignition, resp. during the test
phase of the operating module can
also lead to this error indication. Con-
trol unit or pulse device is defective.
7 segments and the speed/driving
profi le display are "lit":
The joystick communication is defec-
tive. Check the connector cable.
8 segments blink: Defective pulse
device or a system fault. Check cables
and connecting plugs.
8 segments and actuators are
"lit":
One actuator is defective. On two ac-
tuators check which one is not func-
tioning and check the cables on this
one.
9 segments blinks: Fault on the
magnetic brakes of the motor. This
fault indication appears also if the le-
ver of the drive motors is in the push
mode position. Move the lever to the
drive mode position.
10 segments blinks: The battery
voltage is too high (downhill driving)
or the electrical connections are not
O.K. Check the plug on the operating
module and the battery connections.