TP202 20-42-FO-00014 Rev A1 C2169
003-1879-99
Swivel
Coupler
Handpiece
w/Swivel
Handpiece
w/o Swivel
To Clean Handpieces...
1) Detach handpiece from hose and remove heads from contra angle/straight attachments and if applicable Swivel
Couplers. Remove all burs, prophy cups, etc. from heads and straight attachments. Scrub handpiece with a soft
brush and warm water to remove debris. Rinse and dry.
2) Apply Midmark One Step Cleaner/Conditioner as follows:
(A) Clean/Condition Drive Air Port (Applicable for 002-1277-01 Low-Speed w/o Swivel 0-20,000 rpm).
• Assemble spray tube onto spray can head by firmly pushing tube into conditioner can spray tip as far
as possible.
• Insert tube into drive air port of handpiece.
• Direct handpiece away from your body and towards an absorbent towel. Depress handpiece
conditioner can top to spray until conditioner flows from both ends of handpiece head (1 to 2
seconds).
• If overspray on towel appears discolored repeat spray for an additional two seconds. Remove
handpiece and wipe excess conditioner from handpiece with a clean dry towel.
(B) Clean/Condition Swivel (Applicable for 002-1277-00 Low-Speed w/o Swivel 0-5,000 rpm).
• Remove the swivel coupler from the motor by pulling back on the release collar until the swivel pulls
out of the motor connector.
• Clean/Condition the swivel coupler o-rings.
• Clean/Condition the motor swivel ring by applying Midmark One Step Cleaner/Conditioner to the joint
between the swivel ring and the motor.
(C) Operate (Applicable for ALL Low-Speed Handpieces)
• Attach handpiece to hose and run for 8-12 seconds with a bur or prophy angle installed.
• Wipeo󰀨excessconditioner.
3) Seal items in separate sterilization bags. Remove bur or prophy angle from handpiece prior to sealing items in
sterilization bags.
4) Sterilize in steam autoclave as follows:
• Place handpiece(s) in sterilizer tray(s).
• Sterilize at a temperature of 275 degrees Fahrenheit (135 degrees Celsius) for at least 20 minutes.
5) Immediately remove handpiece from autoclave or chemiclave. Allow to cool sufficiently prior to handling.
Midmark Low-Speed Handpiece
Cleaning/Maintenance continued...
Drive Air Port -
Apply Cleaner/
Conditioner Here