Preva Dental X-ray System User Manual, 00-02-1576, Revision ZA1
11
Compliance with Applicable
S
t
a
n
d
ar
d
s
The following regulatory documents
a
pp
ly
Radiation
Protection
The certified components of the Preva Dental X-ray System comply with Radiation
Performance Standards 21 CFR, Subchapter J, at the time of manufacture.
The certified components of the Preva Dental X-
ray System comply with IEC
60601-1-3 Radiation protection/X-ray equipment.
UL 2601-1
File Number:
E181750
Classified by Underwriters Laboratories Inc. with respect to electrical shock, fire
and mechanical hazards only in accordance with UL 2601-1, and CAN/CSA C22.2
NO, 601.1-M90, and to the following particular standards, IEC 60601-2-7, IEC
60601-2-28.
General
Safety
IEC 60601-1:1995
Medical device risk classification Class II (FDA)
Protection against electrical shock Class I (IEC 60601)
Degree of protection against electrical shock
Type B applied partPass (UL)
Type BF applied part (ClearVision sensor only) Pass (UL)
Degree of protection against ingress of water (ClearVision sensor only)IP67
Not suitable for use in the presence of flammable anesthetic mixture with air or
with oxygen or nitrous oxide.
EMI/EMC
IEC 60601-1-2:2007
Degree of
Protection
IEC 60529:2001
Degree of protection against ingress of water (ClearVision sensor only)IP67
Imaging
Performance
IEC 61223-3-4:2000
Line pair resolution better than 8 lp/mm
Low contrast resolution all holes visible