Preva Dental X-ray System User Manual, 00-02-1576, Revision ZA1
38
Specifications
Parameter
Description
Line Voltage
AC 110 V to 230 V, 50 Hz or 60 Hz
Line Load
250 V, UL Recognized It is recommended that branch circuit does
not exceed 15A
Tube Potential
60 kV, 65 kV, 70 kV
Tube Current
4 mA, 5 mA, 6 mA, 7 mA (7 mA is not available at 70 kV)
Irradiation Time
20 ms through 2 s
Maximum Deviation from
Indicated Values
Peak tube potential, maximum deviation: ±5%
Tube current, maximum deviation: ±1 mA
Exposure time: min 20 ms, max. 2 s, max. deviation: ±5%+1 ms
Minimum Source-to-Skin
Distance
20 cm (8 in.)
30 cm (12 in.)
Focal Spot
0.4 mm (per IEC 60336)
Temperatures
Operating
+10 °C to +35 °C (+50 °F to +95 °F)
Storage
-35 °C to +66 °C (-31 °F to +151 °F)
Transport
0 °C to +50 °C (+32 °F to +122 °F)
Atmospheric Pressures
Operating
70 kPa to 106 kPa
Storage
70 kPa to 106 kPa
Transport
70 kPa to 106 kPa
Maximum Altitude
3,000 m (9,843 ft.)
X-ray Beam Dimension
Diameter of 6 cm (2.36 in.) at the end of the 8-inch cone.
Cones with smaller diameter or rectangular beams are available.
Humidity Range
(Operation & Storage)
10 to 80% non-condensing
U.S. Patents
D470237, D469182, D470589, and 6,837,468