Preva Dental X-ray System User Manual, 00-02-1576, Revision ZA1
13
––
Guidance and manufacturer's declaration - electromagnetic emissions
The Preva Dental X-ray System is intended for use in the electromagnetic environment specified below. The customer or the user of the
Preva Dental X-ray System should assure that it is used in such an environment.
Electromagnetic environment – guidance
CISPR 11
The Preva Dental X-ray System uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in nearby
The Preva Dental X-ray System is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emission
IEC 61000-3-2
flicker emissions
Guidance and manufacturer's declaration - electromagnetic immunity
The Preva Dental X-ray System is intended for use in the electromagnetic environment specified below. The customer or the user of the
Preva Dental X-ray System should assure that it is used in such an environment.
Immunity test IEC 60601 test level
Electromagnetic
environment – guidance
(ESD)
IEC 61000-4-2
± 8 kV air
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If the floors are covered with synthetic
material, the relative humidity should be at
least 30%.
transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
supply lines
± 1 kV for input/
Mains power quality should be that of a
transient/ burst supply lines typical
commercial or hospital environment.
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Voltage dips, interruptions,
and voltage variations on
power supply input lines
IEC 61000-4-11
T
T
0.5 cycle
< 40% U
T
(60% dip in U
T
) for 5 cycles
< 70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(>95% dip in U
T
) for 5 s
Power frequency (50/60 Hz)
magnetic field
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.