For future servicing work, please
document any alterations to the stan
-
dard factory settings.
Enter any alterations in the field
"Value set:_________".
General notes:
–
Switch positions that have not been
allocated a function are indicated in
the middle part of the display by a
bar ( - ).
–
The programming level is called up
via the STOP position f.
– If any changes have been made to
programmes or dispensing systems
these must be documented in a log
book kept with the machine in accor-
dance with the Medical Device Di-
rective (MPG directive 93/42/EWG,
Class IIa) and Health and Safety
Regulations.
1. Setting the dosage concen-
tration and venting the dis-
pensing system
^
Turn the programme selector to the
STOP position f.
^
Switch off the machine.
^
Press and hold ß and 6 at the
same time, and whilst doing so,
switch the machine on with the I-0
button.
The current programme status "P..."
will appear in the display and the p /
6 indicator light will come on.
^
Press 0 twice, "E02" (programming
level 2) will appear in the display.
1.1 DOS 1 (liquid cleaning agent) via
the optional DOS-Module G 60 or
C 60.
1.1.1 Setting the dosage concentra
-
tion:
^
Set the dosage selector on the
DOS-Module to position 10 (applies
to DOS Module C 60 only).
Set the dosage concentration accord
-
ing to the cleaning agent manufac
-
turer’s recommendations.
Example: If the cleaning agent manu
-
facturer recommends a concentration
level of 0.5% set the value to 0.50.
^ Turn the programme selector to B.
The display will show either the dos-
age time in seconds for cleaning
agent dispensed via the compart-
ment for powder detergent in the
door, or the dosage concentration for
liquid cleaning agent as a %:
10 = Dosage via the door compart-
ment (10 sec) or
0.10 - 1.0 = 0.1 % -1 % dosage con
-
centration.
^
Press ß as many times as neces
-
sary or hold it pressed in until the re
-
quired value appears.
The dosage time / dosage concentra
-
tion is now set.
Value set:
Programming special functions
39