A.8.5 PR Specifications
PR from Mindray SpO
2
Module
PR from Masimo SpO
2
Module
PR from Nellcor SpO
2
Module
PR from IBP Module
A.8.6 NIBP Specifications
Measurement range 20 to 300 bpm
Resolution 1 bpm
Response time <30 s (normal perfusion, no disturbance, PR value sudden changes from 25
to 220bpm)
Accuracy ±3 bpm
Refreshing rate ≤1 s
Measurement range 25 to 240 bpm
Resolution 1 bpm
Response time ≤20 s (with normal perfusion, no disturbance, and a PR value transition from
25 to 220 bpm)
Accuracy ±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate ≤1 s
Measurement range 20 to 300 bpm
Resolution 1 bpm
Response time ≤30 s (normal perfusion, no disturbance, PR value sudden change from 25
to 250 bpm)
Accuracy 20 to 250 bpm: ±3 bpm
251 to 300 bpm: not specified
Refreshing rate ≤1 s
Measurement range 25 to 350 bpm
Resolution 1 bpm
Accuracy ±1 bpm or ±1%, whichever is greater
Standards Meet standard of IEC 80601-2-30
Technique Oscillometry
Mode of operation Manual, Auto, STAT, Sequence
Auto mode repetition intervals 1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 1.5 h,
2 h, 3 h, 4 h, 8 h
STAT mode cycle time 5 min
Maximum measurement time Adult, pediatric:
Neonate:
180s
90s