A.8.7 Temp Specifications
A.8.8 IBP Specifications
Measurement range Measurement Item Adult Pediatric Neonate
Systolic (mmHg) 25 to 290 25 to 240 25 to 140
Diastolic (mmHg) 10 to 250 10 to 200 10 to 115
Mean (mmHg) 15 to 260 15 to 215 15 to 125
Measurement accuracy* Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Static pressure measurement range 0mmHg to 300mmHg
Static pressure measurement accuracy ±3mmHg
Resolution 1 mmHg
Software overpressure protection Adult:
Pediatric:
Neonate:
297±3 mmHg
297±3 mmHg
147±3 mmHg
Initial cuff inflation pressure range Adult:
Pediatric:
Neonate:
80 to 280 mmHg
80 to 210 mmHg
60 to 140 mmHg
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured with this
device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2)in terms of mean error
and stardard deviation by comparing with intra-arterial or auscultatory measurements (depending on the configuration) in
a typical patient population. For auscultatory reference, the 5th Korotkoff sound was used to determine the diastolic
pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the American
National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and standard deviation by
comparing with intra-arterial measurements (depending on the configuration) in a typical patient population.
PR
Measurement range 30 to300 bpm
Resolution 1 bpm
Accuracy ±3bpm or ±3%, whichever is greater
Standard Meet the standard of ISO 80601-2-56
Technique Thermal resistance
Operating mode Direct mode
Measurement range 0 °C to 50 °C (32 °C to 122 °F)
Accuracy ±0.1 °C or ±0.2 °F (without considering probe error)
Resolution 0.1 °C
Refreshing rate ≤1 s
Minimum time for accurate measurement Body surface: <100 s
Body cavity: <80 s
Standard Meet the standard of IEC 60601-2-34
IBP
Measurement range -50 mmHg to 360mmHg
Resolution 1 mmHg