3.7 Turning Off the Equipment
• Press and hold the power switch for 10s to forcibly shut down the equipment if it could not be shut
down normally. This may cause loss of patient data.
• To prevent the changes from losing in case of sudden power failure, the equipment saves the
settings in real time. In case of a temporary power failure, if the power is restored within 60s, the
equipment will resume with all active settings unchanged; if the power is interrupted for more than
120s, the equipment behaves the same as it is normally turned off; if the power is restored within
60s to 120s, the equipment will resume with all active settings unchanged, or behaves the same as it
is normally turned off.