A - 4
A.2.3 Charger Station Specifications
Battery run time Operating
condition
One battery Two batteries Testing condition
Monitor
mode
≥6.5 h ≥13 h • New fully charged battery.
•Ambient temperature of
20ºC±5 °C.
• The equipment is configured
with 3-/5-lead ECG, manual
defibrillation, screen
brightness set to the lowest
level without printing.
≥5.5 h ≥11 h • New fully charged battery.
•Ambient temperature of
20ºC±5 °C.
• The equipment is configured
with 3-/5-lead ECG, Resp,
SpO2, NIBP measurements set
at an interval of 15 minutes,
manual defibrillation, screen
brightness set to the factory
default without printing.
Defibrillation
mode
≥220 discharges
of 360J
≥440 discharges
of 360J
• New fully charged battery.
•Ambient temperature of
20ºC±5 °C.
• The equipment is configured
with 3-/5-lead ECG, manual
defibrillation, screen
brightness set to the factory
default without printing.
• 3 discharges every minute.
≥300 discharges
of 200J
≥600 discharges
of 200J
Pacer mode ≥4.5 h ≥9 h • New fully charged battery.
•Ambient temperature of
20ºC±5 °C.
• The equipment is configured
with 3-/5-lead ECG, manual
defibrillation, external pacing,
screen brightness set to the
factory default without
printing.
• 50 load impedance, pacer
rate at 80bpm, pacer output at
60mA, pacer pulse at 40ms.
Low alarm
presented
At least 20
minutes of
monitoring and
6 discharges of
360J
At least 40
minutes of
monitoring and
12 discharges of
360J
•Ambient temperature of
20ºC±5 °C.
• The equipment is configured
with 3-/5-lead ECG, manual
defibrillation, screen
brightness set to the factory
default without printing.
Battery fuel gauge 5 LEDs indicating the current battery charge level
Shutdown delay At least 20 minutes of monitoring and 6 discharges of 360J
Maximum number of batteries charged Two batteries can be charged at the same time.
AC power input
Input voltage 100 to 240 VAC (±10%)
Input current 1.8 to 0.8A
Frequency 50/60Hz (±3Hz)