A - 6
A.4.5 External Connectors
A.4.6 Signal Outputs
Beep Gives reminder tones
Audio signal Alarm tone: ISO mode with frequency of 600 Hz
QRS tone: short beep with frequency of 495 Hz
Charge tone: long beep with frequency of 400 Hz to 533 Hz
Charge done tone: double beeps with frequency of 440 Hz
Key tone: short beep with frequency of 1000 Hz
Power input 1, AC power input with equipotential grounding terminal or DC power
input, connects the external power supply.
Multifunctional connector 1, connects a CPR sensor, or a cable for analog output or synchronized
cardioversion.
USB connector • 1 USB 2.0, connects USB devices.
• 1 USB 3.0,
Network connector 1 RJ45 connector, 100 Base-TX, IEEE 802.3, connects a standard network
cable.
Multifunctional connector
Standard Meets the requirements of EN60601-1 for short-circuit protection and
leakage current
ECG Analog Output (only from ECG accessories)
Bandwidth
(-3 dB; reference frequency: 10 Hz)
Diagnostic mode: 0.05 to 150 Hz
Monitor mode: 0.5 to 40 Hz
Therapy mode: 1 to 20 Hz
ST mode: 0.05 to 40 Hz
Maximum QRS delay 25 ms (in diagnostic mode, and with Notch off)
Sensitivity 1 V/mV ±5%
Pace enhancement Signal amplitude: V
oh
≥2.5V
Pulse width: 10ms±5%
Signal rising and falling time: ≤100s
IBP analog output
Bandwidth (-3dB; reference
frequency:1Hz)
0 to 40 Hz
Maximum transmission delay 30 ms
Gain (reference frequency 1 Hz) 1 V/100 mmHg, ±5%
Video output
Video signals Visual graphics captured through a camera.
Synchronous input
Input limit Vih≥2.4v; Vil≤ 0.4v
Input signal range 0 to 5V (TTL level)
Input impedance ≥10 k
Pulse width >5 ms
Alarm output