A - 13
• The equipment startup time in the fast startup mode is less than 2s.
A.7.2 CPR Compression Specifications
Compressions from CPR Sensor
Compressions from Electrode Pads
A.7.3 Pacer Specifications
Power supply
Manual Defib AED
Charge time* From initial
power on
(from cold
start) to charge
done
From initial
power on (from
fast startup
mode) to charge
done
From initiation
of rhythm
analysis to
charge done
From initial
power on
(from cold
start) to charge
done
From initial
power on (from
fast startup
mode) to charge
done
200J 360J 200J 360J 200J 360J 200J 360J 200J 360J 200J 360J
With a new fully
charged battery
<3 s <7 s <11 s <14 s <6 s <10 s <10 s <12 s <21 s <26 s <13 s <15 s
With a new fully
charged battery,
depleted by 15
discharges of 360 J
<4 s <8 s <12 s <15 s <7 s <11 s <11 s <13 s <23 s <27 s <14 s <16 s
With an external
power supply
<4 s <7 s <11 s <14 s <7 s <10 s <11 s <12 s <22 s <24 s <14 s <15 s
* The battery is charged at ambient temperature of 20 ±5 °C.
Compression rate Measurement range: 40 to 160 cpm
Accuracy: ±2 cpm
Compression rate Measurement range: 60 to 200 cpm
Accuracy: ±3 cpm
Standards Meet standards of IEC 60601-2-4
Pacer mode Demand, fixed
Output waveform Monophasic square wave pulse
pulse width 20 ms or 40 ms
Accuracy: ±5%
Pacing rate 30ppm to 210ppm
Accuracy: ±1.5%
Resolution: 5 ppm
Pacing output 0mA to 200mA,
Accuracy: ±5% or ±5mA, whichever is greater
Resolution: 1mA, 2mA or 5mA
Refractory period 200 to 300 ms (depending on pacing rate)
4:1 pacing Pacing pulse frequency reduced by factor of 4 when this function is
activated.