3 - 2
• Do not touch the exposed pins of the transport dock and equipment. Damaged pins affect the
product performance.
• Ensure that the external power system has secure protective earth when the equipment is used
together with the transport dock.
• To avoid a pinch hazard, be careful to connect the transport dock.
• Use the equipment and transport dock on a stable surface.
• Do not stack the equipment connected with a transport dock with other equipments.
• Keep the transport dock away from liquids. Use the transport dock in dry circumstances.
• Do not disassemble, puncture, or incinerate the transport dock.
• When connected with the transport dock, it is specified as a part of the equipment. Use only the
specified transport dock.
3.3 Turning on the Equipment
• Do not use the equipment on a patient if you suspect it is not working properly, or if it is
mechanically damaged. Contact the service personnel or Mindray.
• Check that visual and auditory alarm signals are presented correctly when the equipment is turned
on.
3.4 General Operations
• Check that the touchscreen is not damaged or broken. If there is any sign of damage, stop using the
equipment and contact the service personnel.
• If the touchscreen is loose, stop using the equipment and contact the service personnel.
3.5 Setting Up the Equipment
• Changing the date and time affects the storage of trends and events and may result in loss of data.
• If Brightness is set to Auto, the screen brightness automatically changes according to the ambient
light level.
3.6 Taking Rescue Records
• In the Pacer mode, the equipment automatically takes records for the pacing related operation.
These operations cannot be manually recorded.