EasyManua.ls Logo

Mindray BeneHeart D60 - Measuring Mainstream CO; Zeroing the Mainstream CO Sensor; Changing CO Settings; Setting the Gas Compensation

Mindray BeneHeart D60
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18 - 2
18.3 Measuring Mainstream CO
2
WARNING
When measuring CO2 using the mainstream CO2 module on patients who are receiving or have
recently received anesthetics, connect the gas outlet to the anesthesia machine or ventilator. This
avoids exposing medical staff to anesthetics.
NOTE
Be sure to set the barometric pressure properly before using the mainstream CO
2
module. Improper
settings will result in erroneous CO
2
reading.
Always position the sensor with the adapter in an upright position to avoid collection of fluids in the
windows of the adapter. Large concentrations of fluids at this point will obstruct gas analysis.
To avoid dead space, place the sensor as close to the patient as possible.
18.3.1 Zeroing the Mainstream CO
2
Sensor
WARNING
When perform a zero calibration during the measurement, disconnect the sensor from the patient’s
airway first.
Do not rely on the readings during CO
2
zeroing.
18.4 Changing CO
2
Settings
18.4.1 Setting the Gas Compensation
WARNING
Make sure to use the appropriate compensations. Inappropriate compensations may cause
inaccurate measurement values and result in misdiagnosis.
18.4.2 Changing Barometric Pressure
WARNING
Be sure to set the barometric pressure properly before using the mainstream CO
2
module. Improper
settings will result in erroneous CO
2
reading.
18.5 CO
2
Calibration
CAUTION
Connect the gas outlet to the scavenging system when calibrating the CO
2
module.
18.6 CO
2
Troubleshooting
NOTE
For the physiological and technical alarm messages, see DAlarm Messages.

Table of Contents

Related product manuals