A.6.5 NFC Specifications
A.7 Therapy Specifications
A.7.1 Defibrillation Specifications
Test conditions Meets the following conditions simultaneously:
• With a distance greater than 2m from each other, three equipments
operate properly at the same time.
• Bluetooth communication does not interrupt.
• The distance between the interfering devices and the equipment is
greater than 20 cm. A Wi-Fi interference (no greater than -85 dBm) in the
same channel and a Wi-Fi interference (no greater than -50 dBm) in an
adjacent-channel are presented synchronously. The interfering devices
include, but are not limited to, 2.4 G wireless devices, cellular mobile
networks, interphones, cordless phones, and ESU equipment. The
interfering devices do not include Wi-Fi devices.
Protocol ISO/IEC 14443 A, ISO/IEC 1443 B
Operating frequency 13.56 MHz
Standards Meet standards of IEC 60601-2-4
Defibrillation mode Manual defib, synchronized cardioversion, AED
Defibrillation waveform Biphasic truncated exponential (BTE) waveform, auto-compensation
according to patient impedance
Defibrillation electrodes External paddles set coming with pediatric paddles included, multifunction
electrode pads and internal paddles
Controls and indicators on external
paddles
Charge button, Shock buttons, Energy Select buttons, charge done indicator
and patient contact indicator
Controls and indicators on internal
paddles
Shock button
Shockable rhythm analysis time < 5 s
Range of selected energy
External defibrillation 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, 50, 70, 100, 120, 150, 170, 200, 300, 360
J
Internal defibrillation 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 25, 30, 50 J
Patient impedance range
External defibrillation 25 to 300
Internal defibrillation 15 to 300
Synchronized discharge delay
Local synchronized discharge delay < 60ms (from the peak of R-wave)
Remote synchronized discharge delay < 25ms (from the rise edge of synchronous signal)
AED
Shock series Energy level: 100 to 360J, configurable for adult;
10 to 100J, configurable for pediatric
Shocks: 1, 2, 3, configurable;
Meeting AHA/ERC guidelines 2015 by default
AED ECG analysis performance Refer to CMindray Shockable Rhythm Analysis Algorithm.