ELECTRICAL SAFETY INSPECTION FORM
American version
Overall assessment:
Scheduled inspection Test item: 1, 2, 3, 9
Unopened repair type Test item: 1, 2, 3, 9
Opened repair type, not modify the power part
including transformer or patient circuit board
Test item: 1, 2, 3, 4, 5, 9
Opened repair type, modify the power part
including transformer or patient circuit board
Test item: 1, 2, 3, 4, 5, 6, 7, 8, 9
Location Technician
Equipment Control Number
Manufacturer Model SN
Measurement equipment /SN Date of Calibration
INSPECTION AND TESTING
Pass/Fail Comments
1 Power Cord Plug
2 Device Enclosure and Accessories
3 Device Labeling
4 Protective Earth Resistance Limit 0.2
5 Earth Leakage
Limit
NC:300μA
SFC:1000μA
Normal condition(NC) μA
Single Fault condition(SFC) μA
6* Patient Leakage Current
Limit
NC:10μA,
SFC: 50μA
Normal condition(NC) μA
Single Fault condition(SFC) μA
7* Mains on Applied Part Leakage Limit 50μA
8* Patient Auxiliary Current
Limit
NC:10μA,
SFC: 50μA
Normal condition(NC)
Single Fault condition(SFC)
9 Functional test (parameters tested):
Note: The test items marked “*” are needed only for incoming inspections and after repairs or
modifications that may have affected lead leakage [NFPA 99 (2005)8.5.2.1.3].
Deficiency / Note:
Name: _____________________________ Date / Signature: _________________________________

Table of Contents

Other manuals for Mindray DP-9900

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DP-9900 and is the answer not in the manual?

Mindray DP-9900 Specifications

General IconGeneral
BrandMindray
ModelDP-9900
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals