S-5 
    (6)  Confirm that the peripheral units are secured before you 
move the system. Otherwise, the peripheral units may fall 
and cause injury. 
    (7)  When the system is moved over a step, exercise caution to 
protect the system from falling. When holding the system at 
the bottom to help move it over a step, take special care to 
prevent hand injuries. 
 4.Do not vibrate the equipment excessively (when moving the 
equipment); otherwise the mechanical parts (such as casters) may 
be damaged. If the system is often moved on a bumpy floor, contact 
the Mindray sales office, customer service department or 
representative. 
  5.  Do not connect this system to outlets with the same circuit 
breakers and fuses that control the current of devices such as 
life-support systems. If this system malfunctions and generates an 
overcurrent, or when there is an instantaneous current at power 
ON, the circuit breakers and fuses of the building’s supply circuit 
may be tripped. 
 6. Always keep the machine dry. Avoid transporting this machine 
quickly from the cold place to the warm place, otherwise 
condensation or water drops may be formed, which will cause short 
circuit. 
 7. If the circuit breaker is tripped or the fuse is blown, it indicates that 
the machine or the peripheral devices have problems. In these 
cases, the user cannot repair by him but contact the Mindray sales 
office, customer service department or representative. 
 8. There is no risk of high-temperature burns during routine 
ultrasound examinations, even if, due to environment temperature 
and exam modes, the surface temperature of the transducer exceeds 
the body temperature of the patient. To prevent high-temperature 
burns, do not apply the transducer to the same region on the patient 
for a long time. Apply the transducer only for as long as required 
time for diagnosis.  
 9.After the sterilization or disinfection of accessories, chemicals must 
be washed out or gases must be discharged thoroughly from the 
accessories. Remaining residual chemicals or gases will not only 
result in damage to the accessories but also can be harmful to 
human bodies.