EasyManuals Logo

Mindray Eagus R9 User Manual

Mindray Eagus R9
365 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #195 background imageLoading...
Page #195 background image
Structure and Assembly/Disassembly 9-33
6. Remove the absorbing cotton insulation.
Note:
Do not tilt or block the screw holes when installing the absorbing cotton insulation
anticlockwise.
7. Unscrew 4 M4 X 12 cross panhead screws with screwdriver (M3, M4) to remove 4 clips.
Note:
Be aware of the orientation of the clip. Align the locating pin of the clip with the locating
hole of the control panel base when performing the installation.
8. Unscrew 6 M5 X 16 round inner head screw (with the pad) by M5 spanner to remove the base
of the side control panel upwards.