16 Probes and Biopsy
Operator’s Manual 16 - 21
b. Screw the block's nut to secure the block.
c. Insert a biopsy needle with the same specification as that of the guiding block into the
guiding block hole.
4. Release the needle from the bracket:
a. Screw the nut of the guiding block and remove the guiding block slightly along the
direction of the needle’s tail.
b. Separate the residual part of the needle-guide bracket and the probe from the needle.
5. Remove the needle-guided bracket:
a. Unscrew the Needle-guided bracket pinch nut, and remove the needle-guided bracket
from the probe.
b. Separate the probe and the needle-guided bracket.