18 - 4 Operator’s Manual
18 System Maintenance
Cleaning the dust-proof covers
CAUTION
Please clean all dust-proof covers of the system periodically (1 time per
month); otherwise, system damage may result. Cleaning times can be
increased when the system is used in the open air or somewhere dust is more.
Tool: soft brush
Method:
Disassemble dust-proof cover before cleaning.
18.1.2 Disinfecting the Main Unit
Only the system equipped with trackpad supports disinfection.
WARNING
Use only Mindray approved disinfectants and methods listed in this section
to disinfect the main unit. Warranty does not cover damage caused by
unapproved substances or methods.
Do not mix disinfectants, as hazardous gases may result.
We make no claims regarding the efficacy of the listed chemicals or
methods as a means for controlling infection. For the method to control
infection, consult your hospital’s infection control officer or epidemiologist.
CAUTION
Never immerse any part of the main unit in liquids or allow liquid to enter
the interior.
Any contact of disinfectants with connectors or metal parts may cause
corrosion.
Do not pour or spray any liquid directly on the main unit or permit fluid to
seep into connections or openings.